FH635
Phone: 1-800-757-3747 | Fax: 1-800-676-0498 | Email: dicksoncsr@dicksondata.com | For calls outside the US please call: 1-630-543-3747
_________________________________________________________________________________________________
Touchscreen
_________________________________________________________________________________________________
Specifications
Alarm Type: Audio/Visual
Ambient Operating RH Conditions : 0 to 95% RH non-condensing (Logger Only)
Ambient Operating Temperature Conditions: 0 to 122F (-17 to 50C)
Battery Backup: 12 Hrs (Re-chargeable batteries included; Hardwired)
Cable Details: USB - Male Series A plug to 5 pin Male Series B mini plug.
Calibration Options: N100 , N300, N400, & N520
Channels: CH1: Temperature (F/C), CH2: Humidity
Data Capacity: 32,000 (16,000 per channel)
Dew Point Accuracy: +/-3.6F from 30 to 95% RH (Calculated) (+/-2C from 30 to 95% RH (Calculated))
Dimensions: 1.4 x 8.5 x 6.9in (3.5 x 21.6 x 17.5cm)
Display Dimensions Height: 5
Display Dimensions Width: 6.4
Display Type: Touchscreen
Display Update Rate: Varies with logging interval
Displays Min/Max: Yes
Download Type: USB / Flash Card
Flash Card Option: SD Card (2GB or less)
Humidity Accuracy: +/-2% RH from 0 to 60%; +/-3% RH from 60 to 95%
Humidity Range: 0-95% RH non-condensing
IP Rating: 21
Key Features: Touchscreen, multiple screen views, on-screen zoom and scroll, software optional, user customizable
sample rate, temperature scales and unit of measure
Minimum Software Version Required: 17.2.0
Mounting Options: Mounting Bracket (Included)
Power Cord: 6 ft
Power Source: 120 to 240 VAC, 12 VDC (INTL adapters available)
Relay Type: SPST, 24 VDC. 0.5A, N/O N/C
Remote Probe Dimensions : Houses thermistor and RH sensor in polycarbonate plastic; 2.12" x 1.2" x .5"
Remote Sensor Length: Single 10 ft extension cable available (without sacrificing accuracy)
Remote Sensor Type: 1 Field Replaceable Digital Temperature/RH Sensor
Required Software: A016, A026, N525 (all optional)
Response Time: 1 Minute to 63% of Full Scale
Sample Interval: User Selectable via Time Scale (See Product Manual for more details)
Temperature Accuracy: +/-0.8F from 20 to 122F 1.8F Remaining (+/-0.45C from -6 to 50C 1C Remaining)
Temperature Range: -40 to 185F (-40 to 85C)
Unit Weight: 1.15 lbs
User Calibration with SW400: Yes
Warranty: 12-Month Limited
Vue de la page 0
1 2 3 4 5 6 ... 11 12

Résumé du contenu

Page 1 - Touchscreen

FH635Phone: 1-800-757-3747 | Fax: 1-800-676-0498 | Email: [email protected] | For calls outside the US please call: 1-630-543-3747___________

Page 2

FH635Phone: 1-800-757-3747 | Fax: 1-800-676-0498 | Email: [email protected] | For calls outside the US please call: 1-630-543-3747___________

Page 4

AC Adapter Extension Cable10' Power Extension for AC Adapter_____________________________________________________________________________________

Page 5

FH635Phone: 1-800-757-3747 | Fax: 1-800-676-0498 | Email: [email protected] | For calls outside the US please call: 1-630-543-3747___________

Page 6

FH635Phone: 1-800-757-3747 | Fax: 1-800-676-0498 | Email: [email protected] | For calls outside the US please call: 1-630-543-3747___________

Page 8

Memory CardRemovable Flash Memory CardCertificate of ValidationCertificate of ValidationCard Reader KitIncludes USB Reader and Flash Memory Card

Page 10 - Accessories FH635

Sensaphone DialerRemote Monitoring Unit - calls 4 numbers

Page 12 - Calibrations

Sensaphone DialerRemote Monitoring Unit - calls 8 numbersUniversal Plug Kit for AC AdapterSet of Three International Power Plugs for AC Adapter

Commentaires sur ces manuels

Pas de commentaire